Master Vacuum Pumps
Brand(s)
Last updated
Summary
Product
Master Vacuum Pumps
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Master Vacuum Pumps | 2315316 2315317 | 30.00.30 30.00.35 17-8689 |
Issue
When the push-start board was manufactured, a capacitor was incorrectly mounted (polarity). This can lead to overheating if the device is running for min. 60 minutes and then restarted. Delay for restart and treatment.
Recall start date: November 6, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Ardo medical AG
Gewerbestrasse 19, Unterageri, Switzerland, 6314
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74621
Get notified