Health product recall

Master Vacuum Pumps

Brand(s)
Last updated

Summary

Product
Master Vacuum Pumps
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Master Vacuum Pumps

2315316
2315317

30.00.30
30.00.35
17-8689

Issue

When the push-start board was manufactured, a capacitor was incorrectly mounted (polarity). This can lead to overheating if the device is running for min. 60 minutes and then restarted. Delay for restart and treatment. 

Recall start date: November 6, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Ardo medical AG

Gewerbestrasse 19, Unterageri, Switzerland, 6314

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74621

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe