MassCheck Cortisol, Cortisone Saliva Control Level 1 & II (2019-09-07)
- Starting date:
- September 7, 2019
- Posting date:
- November 15, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71593
Last updated: 2019-11-15
Affected Products
- MassCheck Cortisol, Cortisone Saliva Control Level 1
-
MassCheck Cortisol, Cortisone Saliva Control Level II
Reason
The batches on the market of order no 730390, 0349 and 0350 show a time-dependent stability related deviation from the target values. Using the affected calibrator batches may result in increased results of patient samples, with a positive bias of approx. 15 % detectable. However the measured concentrations are still within the Chromsystems internal acceptance criteria.
Affected products
-
MassCheck Cortisol, Cortisone Saliva Control Level 1
Lot or serial number
1517
Model or catalog number
0349
Companies
- Manufacturer
-
Chromsystems Instruments & Chemicals GMBH
Am Haag 12
Grafelfing
82166
GERMANY
B. MassCheck Cortisol, Cortisone Saliva Control Level II
Lot or serial number
1517
Model or catalog number
0350
Companies
- Manufacturer
-
Chromsystems Instruments & Chemicals GMBH
Am Haag 12
Grafelfing
82166
GERMANY