Health product recall

MassCheck Cortisol, Cortisone Saliva Control Level 1 & II (2019-09-07)

Starting date:
September 7, 2019
Posting date:
November 15, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71593


Last updated:

Affected Products

  1. MassCheck Cortisol, Cortisone Saliva Control Level 1
  2. MassCheck Cortisol, Cortisone Saliva Control Level II

    Reason

    The batches on the market of order no 730390, 0349 and 0350 show a time-dependent stability ­related deviation from the target values. Using the affected calibrator batches may result in increased results of patient samples, with a positive bias of approx. 15 % detectable. However the measured concentrations are still within the Chromsystems internal acceptance criteria.

    Affected products

    1. MassCheck Cortisol, Cortisone Saliva Control Level 1

    Lot or serial number

    1517

    Model or catalog number

    0349

    Companies
    Manufacturer

    Chromsystems Instruments & Chemicals GMBH

    Am Haag 12

    Grafelfing

    82166

    GERMANY



    B. MassCheck Cortisol, Cortisone Saliva Control Level II

    Lot or serial number

    1517

    Model or catalog number

    0350

    Companies
    Manufacturer

    Chromsystems Instruments & Chemicals GMBH

    Am Haag 12

    Grafelfing

    82166

    GERMANY