Health product recall

Martin Lewis Agar w/ LCAT (2018-12-03)

Starting date:
December 3, 2018
Posting date:
January 23, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68890

Affected products

Reason

May not provide acceptable microbiological performance over time. Although it passed the initial Quality Control (QC), decreased to no growth of ATCC strains were observed in product stored at 2 to 12°C.

Affected products

Martin Lewis Agar w/ LCAT

Lot or serial number

2424095

Model or catalog number

MP1337

Companies
Manufacturer
Oxoid Limited
Wade Road
Basingstoke, Hampshire
RG24 8PW
UNITED KINGDOM