Marquee Disposable Core Biopsy Instrument Kit
Brand(s)
Last updated
Summary
Product
Marquee Disposable Core Biopsy Instrument Kit
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Marquee Disposable Core Biopsy Instrument | More than 10 numbers, contact manufacturer. | MQ1210 MQ1213 |
Marquee Disposable Core Biopsy Instrument Kit | More than 10 numbers, contact manufacturer. | MQK1210 MQK1213 |
Issue
Based on customer complaints, Bard Peripheral Vascular, Inc. has identified that specific lots of the Marquee™ Disposable Core Biopsy Instruments may have been potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device.
Recall Start Date: September 30, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies
Bard Peripheral Vascular, Inc.
1625 West 3rd Street, Tempe, Arizona, United States, 85281
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76311
Get notified
Receive notifications for new and updated recalls and alerts by category.