Maquet Hemoconcentrator Set With Softline Coating
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Hemoconcentrator Set With Softline Coating |
3000382838 |
BO-H 48773 |
Issue
An issue with eight technical drawings, specifically the absence of the bonding symbol for certain connections on the tubing set was found. It is important to note that the specified bonding procedure, which involved the use of cyclohexanone, was not carried out. While the absence of the bonding on the connectors may not be immediately apparent, it could potentially lead to leakage in the affected connections.
Recall start date: December 9. 2024
Additional information
Details
Maquet Cardiopulmonary GmbH
Kehler Strasse 31, Rastatt, Germany, 76437