This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Maquet Critical Care Servos U and N (2014-04-17)

Starting date:
April 17, 2014
Posting date:
May 8, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39385

Recalled Products

A. Servo-U

B. Servo-N

Reason

We have had one occurrence of a disconnection from the ventilator tubing, which was not detected by the ventilator, (i.e. No alarms were triggered.) This absence of alarms may occur with system version 1.00.05 in "invasive ventilation modes", in "adult and pediatric patient categories" and with the low peep alarm turned off by setting it to zero. This absence of alarms does not occur in the neonatal patient category, or in non-invasive modes. This absence of alarms does not occur in system version 1.00.06, which is readily available.

Affected products

A. Servo-U

Lot or serial number

20061

20065

20085

20087

20088

20127

 

Model or catalog number

6694800

 

Companies
Manufacturer

Maquet Critical Care AB

Rontgenvagen 2,

Solna

17154

SWEDEN



B. Servo-N

Lot or serial number

1007

1015

 

Model or catalog number

6688600

 

Companies
Manufacturer

Maquet Critical Care AB

Rontgenvagen 2,

Solna

17154

SWEDEN