Maquet Critical Care Servos U and N (2014-04-17)
- Starting date:
- April 17, 2014
- Posting date:
- May 8, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-39385
Recalled Products
A. Servo-U
B. Servo-N
Reason
We have had one occurrence of a disconnection from the ventilator tubing, which was not detected by the ventilator, (i.e. No alarms were triggered.) This absence of alarms may occur with system version 1.00.05 in "invasive ventilation modes", in "adult and pediatric patient categories" and with the low peep alarm turned off by setting it to zero. This absence of alarms does not occur in the neonatal patient category, or in non-invasive modes. This absence of alarms does not occur in system version 1.00.06, which is readily available.
Affected products
A. Servo-U
Lot or serial number
20061
20065
20085
20087
20088
20127
Model or catalog number
6694800
Companies
- Manufacturer
-
Maquet Critical Care AB
Rontgenvagen 2,
Solna
17154
SWEDEN
B. Servo-N
Lot or serial number
1007
1015
Model or catalog number
6688600
Companies
- Manufacturer
-
Maquet Critical Care AB
Rontgenvagen 2,
Solna
17154
SWEDEN