Health product recall

Mammomat Inspiration System - Main Unit (2014-10-29)

Starting date:
October 29, 2014
Posting date:
November 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42097

Recalled Products   

Mammomat Inspiration System - Main Unit 

Reason

Siemens is notifying customers of a potential issue and risk of injury when using the Mammomat inspiration acquisition workstation table when it is not fixed to the floor. If the monitors are positioned too far from the table top, the acquisition workstation table can become unstable with a risk of it falling over and causing injury to the user.

Affected products

Mammomat Inspiration System - Main Unit 

Lot or serial number

6007

Model or catalog number

101 40 000

Companies
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY