Mahurkar Acute Dual Lumen Catheter Kits
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Mahurkar Acute Dual Lumen Catheter Kits |
More than 10 numbers, contact manufacturer. |
8888135161 |
Issue
During the production process, a potential internal leaking condition within the hub of specific Mahurkar acute dual lumen high flow hemodialysis catheter was identified as a result of a void in the catheter hub. During dialysis, this observed adverse internal leaking condition could translate into cross communication of the blood circuit.
Recall start date: December 7, 2022
Additional information
Details
Covidien Llc
15 Hampshire Street, Mansfield, Massachusetts, United States, 02048
Get notified
Receive notifications for new and updated recalls and alerts by category.