MAGNUS 1180 OPERATING TABLE COLUMNS (2021-02-18)

Starting date:
February 18, 2021
Posting date:
March 4, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75069



Last updated:
2021-03-04

Affected Products 

MAGNUS 1180 OPERATING TABLE COLUMNS

Reason

Under certain conditions, they have found that a problem may cause the operating table not to function as intended. The tables may develop a hydraulic leakage at the inclination function. This leakage leads to an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning. This malfunction only exhibits when tilt movement is used. They are aware of instances where this problem led to patients having to be transferred to another OR table which prolonged the procedure.

Affected products

MAGNUS 1180 OPERATING TABLE COLUMNS

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

118001

Companies
Manufacturer

Maquet GmbH

Kehler Strasse 31

Rastatt

76437

GERMANY