MAGNUS 1180 OPERATING TABLE COLUMNS (2021-02-18)
- Starting date:
- February 18, 2021
- Posting date:
- March 4, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75069
Last updated:
2021-03-04
Affected Products
MAGNUS 1180 OPERATING TABLE COLUMNS
Reason
Under certain conditions, they have found that a problem may cause the operating table not to function as intended. The tables may develop a hydraulic leakage at the inclination function. This leakage leads to an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning. This malfunction only exhibits when tilt movement is used. They are aware of instances where this problem led to patients having to be transferred to another OR table which prolonged the procedure.
Affected products
MAGNUS 1180 OPERATING TABLE COLUMNS
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
118001
Companies
- Manufacturer
-
Maquet GmbH
Kehler Strasse 31
Rastatt
76437
GERMANY