Magnetom Free.Max
Last updated
Summary
Product
Magnetom Free.Max
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Magnetom Free.Max |
206217 |
11408425 |
Magnetom Free.Max |
206217 |
11408424 |
Issue
The service pack NX VA60A-SP04 is a bug fix update (full installation) that solves/improves the following:
- Latest it security deliveries and increase in overall stability (syngo MR XA60A-SD12 content included)
- Installation of multiple SW delta packages now possible at the same time
- EU DataAact (EY) 2023/2854 implementation:access to all machine data
- Solved image loading errors for MR Breast Biopsy
- General MR Protocol Manager improvements
- Coil specifications adjusted for BM Body 18 Pilot Tone and Spine 32 and 1.5T Sentinelle Breast coils image brightness
- Supporting IEC 60601-1 compliance, enabling predictable and safe system behavior with regard to table movement
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
| Siemens Shenzhen Magnetic Resonance Ltd. |
| Siemens Mri Center, Shenzhen, China, 518057 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82867
Get notified
Receive emails about new and updated recall and safety alerts.