Health product recall

Magnetom Free.Max

Last updated

Summary

Product
Magnetom Free.Max
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Magnetom Free.Max

206217

11408425

Magnetom Free.Max

206217

11408424

Issue

The service pack NX VA60A-SP04 is a bug fix update (full installation) that solves/improves the following:
 

  • Latest it security deliveries and increase in overall stability (syngo MR XA60A-SD12 content included)
  • Installation of multiple SW delta packages now possible at the same time
  • EU DataAact (EY) 2023/2854 implementation:access to all machine data
  • Solved image loading errors for MR Breast Biopsy
  • General MR Protocol Manager improvements
  • Coil specifications adjusted for BM Body 18 Pilot Tone and Spine 32 and 1.5T Sentinelle Breast coils image brightness
  • Supporting IEC 60601-1 compliance, enabling predictable and safe system behavior with regard to table movement
     

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Siemens Shenzhen Magnetic Resonance Ltd.
Siemens Mri Center, Shenzhen, China, 518057
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82867

Get notified

Receive emails about new and updated recall and safety alerts.