Health product recall

MAESTRO 4000 Cardiac Ablation System (2018-06-20)

Starting date:
June 20, 2018
Posting date:
July 13, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67218

Affected products

MAESTRO 4000 Cardiac Ablation System

Reason

Boston Scientific has received reports of unintended continuation of radiofrequency (RF) energy delivery from the Maestro 4000 Cardiac Ablation System after the foot switch has been released. Boston Scientific has confirmed the Maestro 4000 is meeting performance expectations and that this rare behavior can be resolved with existing instructions within the operator's manual. Unintended RF energy delivery can be detected if, after releasing the foot switch, RF parameters continue displaying on the control panel and the system continues emitting an audible tone. If this behavior occurs after releasing the foot switch, follow instructions within the manual to stop RF delivery by pressing the RF Power Control Button on either the System Controller (Generator) or System Remote Control (Remote).

Affected products

MAESTRO 4000 Cardiac Ablation System

Lot or serial number

1117170669

1117170670

1228170704

412170449

427170482

Model or catalog number

M00440000

M0044000D0

M0044000H0

M0044000R0

Companies
Manufacturer

Boston Scientific Corporation

300 Boston Scientific Way

Marlborough

01752

Massachusetts

UNITED STATES