MAESTRO 4000 Cardiac Ablation System (2018-06-20)
- Starting date:
- June 20, 2018
- Posting date:
- July 13, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67218
Affected products
MAESTRO 4000 Cardiac Ablation System
Reason
Boston Scientific has received reports of unintended continuation of radiofrequency (RF) energy delivery from the Maestro 4000 Cardiac Ablation System after the foot switch has been released. Boston Scientific has confirmed the Maestro 4000 is meeting performance expectations and that this rare behavior can be resolved with existing instructions within the operator's manual. Unintended RF energy delivery can be detected if, after releasing the foot switch, RF parameters continue displaying on the control panel and the system continues emitting an audible tone. If this behavior occurs after releasing the foot switch, follow instructions within the manual to stop RF delivery by pressing the RF Power Control Button on either the System Controller (Generator) or System Remote Control (Remote).
Affected products
MAESTRO 4000 Cardiac Ablation System
Lot or serial number
1117170669
1117170670
1228170704
412170449
427170482
Model or catalog number
M00440000
M0044000D0
M0044000H0
M0044000R0
Companies
- Manufacturer
-
Boston Scientific Corporation
300 Boston Scientific Way
Marlborough
01752
Massachusetts
UNITED STATES