m2000sp Instrument, G-series and E-series (2015-12-14)
- Starting date:
- December 14, 2015
- Posting date:
- January 7, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-56550
Affected Products
A. m2000sp Instrument, G-series and E-series
Reason
Investigation at Abbott Molecular identified that an incorrect type of tip adapter (type D) could have been installed on m2000sp instrument during a service visit on or after august 14, 2015. This is because two lots of Abbott Molecular PN 30084426 (lots 76258007 and 77228007) were shipped with type D tip adapters rather than the required type B adapters.
Affected products
A. m2000sp Instrument, G-series and E-series
Lot or serial number
- 10378
- 10417
- 10472
- 10621
- 10762
- 4123-G SERIES
Model or catalog number
- 9K14-01
- 9K14-02
Companies
- Manufacturer
-
Abbott Molecular Inc.
1300 E. Touhy Ave.
Des Plaines
60018
Illinois
UNITED STATES