Health product recall

m2000sp Instrument, G-series and E-series (2015-12-14)

Starting date:
December 14, 2015
Posting date:
January 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56550

Affected Products

A. m2000sp Instrument, G-series and E-series

Reason

Investigation at Abbott Molecular identified that an incorrect type of tip adapter (type D) could have been installed on m2000sp instrument during a service visit on or after august 14, 2015. This is because two lots of Abbott Molecular PN 30084426 (lots 76258007 and 77228007) were shipped with type D tip adapters rather than the required type B adapters.

Affected products

A. m2000sp Instrument, G-series and E-series

Lot or serial number
  • 10378
  • 10417
  • 10472
  • 10621
  • 10762
  • 4123-G SERIES
Model or catalog number
  • 9K14-01
  • 9K14-02
Companies
Manufacturer

Abbott Molecular Inc.

1300 E. Touhy Ave.

Des Plaines

60018

Illinois

UNITED STATES