This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Luxor / LX3 Microscope Floor Stand Unit

Starting date:
June 1, 2017
Posting date:
June 21, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63690

Reason

Through Alcon's monitoring process, Alcon has observed that multiple shipments or forceful impact to a caster on a Luxor/LX3 microscope may stress the microscope base, which could contribute to a caster breaking or detaching from the base. This may lead to the potential for physical injury from the microscope tipping or falling.

Affected products

Luxor / LX3 Microscope Floor Stand Unit

Lot or serial number

1501088003X

1502672105X

1502857503X

1503218902X

1503261602X

1601025102X

1601093703X

1602750703X

1604011406X

1701277604X

Model or catalog number

8065752560

Companies

Manufacturer

Alcon Laboratories Inc.

6201 South Freeway

Fort Worth

Texas

UNITED STATES