Health product recall

Luminos Agile Max

Last updated

Summary

Product
Luminos Agile Max
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products

Lot or serial number

Model or catalog number

Luminos Agile Max

61694

10762472

Issue

Results of post market surveillance revealed that under certain circumstances the room configuration parameters of the system are set to default values. If these values are larger than the actual room dimensions, a collision with the ceiling or wall cannot be ruled out during system movement. Siemens Healthineers has not become aware of any injuries caused by this issue.

Recall start date: December 16, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Siemens Healthcare Gmbh

Henkestr. 127, Erlangen, Germany, 91052

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-72501

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe