This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Lumex Patriot Homecare Bed

Starting date:
May 14, 2012
Posting date:
July 9, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-26335

Recalled Products

  1. Lumex Patriot Homecare Bed

Reason

Correction and removal is being conducted for the Lumex Patriot Full-Electric Homecare "bed ends". There is potential to provide room for the leg pin to move during use. If failure were to occur, the bed could descend rapidly causing non-typical patient movement and could pose a risk of injury to the patient and others.

Affected products

A. Lumex Patriot Homecare Bed

Lot or serial number

More than 10 Numbers, Contact manufacturer

Model or catalog number
  • 690-7004-000
  • 690-7604-000
Companies
Manufacturer

Basic American Div. Graham Field