Lo-Pro Tracheal Tube Cuffed Murphy Tip
Brand(s)
Last updated
Summary
Product
Lo-Pro Tracheal Tube Cuffed Murphy Tip
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Lo-Pro Tracheal Tube Cuffed Murphy Tip | 22G0344JZX | 86051 |
Issue
Starting in September 2022, customers reported tracheal tube inflation system leaks. Medtronic's investigation identified a manufacturing assembly error which resulted in an incomplete closure of the inflation system.
Recall start date: February 10, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Covidien LLC
15 Hampshire Street, Mansfield, Massachusetts, United States, 02048
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72959
Get notified