Health product recall

Lo-Pro Tracheal Tube Cuffed Murphy Tip

Brand(s)
Last updated

Summary

Product
Lo-Pro Tracheal Tube Cuffed Murphy Tip
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Lo-Pro Tracheal Tube Cuffed Murphy Tip

22G0344JZX

86051

Issue

Starting in September 2022, customers reported tracheal tube inflation system leaks. Medtronic's investigation identified a manufacturing assembly error which resulted in an incomplete closure of the inflation system.

Recall start date: February 10, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Covidien LLC

15 Hampshire Street, Mansfield, Massachusetts, United States, 02048

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72959