Liquid Immunoassay Premium (2019-02-12)
- Starting date:
- February 12, 2019
- Posting date:
- March 8, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69236
Affected Products
Liquid Immunoassay Premium
Reason
The target value for T uptake on the Siemens IMMULITE 2000 is incorrect on the value sheet for the products listed.
The target value for this analyser is listed as 43.8%U and should be 39.4 %U.
Affected products
Liquid Immunoassay Premium
Lot or serial number
1723EC
Model or catalog number
LIA 3105
Companies
- Manufacturer
-
Randox Laboratories Ltd.
55 Diamond Road
Crumlin, County Antrim
BT29 4QY
UNITED KINGDOM