This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Liquicheck Specialty Immunoassay Controls (2016-02-05)

Starting date:
February 5, 2016
Posting date:
February 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57194

Affected Products

  1. LIQUICHEK SPECIALTY IMMUNOASSAY CONTROL
  2. LIQUICHEK SPECIALTY IMMUNOASSAY CONTROL LEVEL LTA
  3. LIQUICHEK SPECIALTY IMMUNOASSAY CONTROL FOUR LEVEL

Reason

Intact PTH may not meet the thawed unopened and thawed open vial stability claims as printed in the package inserts. The claims for intact PTH have been revised.

Affected products

  1. LIQUICHEK SPECIALTY IMMUNOASSAY CONTROL

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

364

365

366

Companies
Manufacturer

Bio-Rad Laboratories

9500 Jeronimo Road

Irvine

92618

California

UNITED STATES



B. LIQUICHEK SPECIALTY IMMUNOASSAY CONTROL LEVEL LTA

Lot or serial number

57431L

57441L

57451L

57461L

57471L

57481L

57491L

Model or catalog number

359

Companies
Manufacturer

Bio-Rad Laboratories

9500 Jeronimo Road

Irvine

92618

California

UNITED STATES



C. LIQUICHEK SPECIALTY IMMUNOASSAY CONTROL FOUR LEVEL

Lot or serial number

57430

57440

57450

57460

57470

57480

57490

Model or catalog number

359X

Companies
Manufacturer

Bio-Rad Laboratories

9500 Jeronimo Road

Irvine

92618

California

UNITED STATES