Health product recall

LIKOGUARD L AND LIKOGUARD XL OVERHEAD LIFTS (2021-07-02)

Starting date:
July 2, 2021
Posting date:
August 6, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-76183



Last updated:
2021-08-06

Affected Products 

LIKOGUARD L AND LIKOGUARD XL OVERHEAD LIFTS

Reason

Hillrom has become aware of a potential issue related to the length of the Emergency strap on overhead lifts LikoGuard L and XL. Through the investigation, Hillrom identified that the Emergency strap could be longer than the intended length after installation of the overhead lift. The Emergency Stop Cord length should be adjusted if needed to an approximate 1400 +/-100 mm above the floor.

Affected products

LIKOGUARD L AND LIKOGUARD XL OVERHEAD LIFTS

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

3301030

3301040

Companies
Manufacturer

Liko AB

Nedre Vagen 100

Lulea

97592

SWEDEN