This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Liko Slingbars, Twinbar and Sidebar (2013-04-30)

Starting date:
April 30, 2013
Posting date:
October 31, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36491

Recalled Products

A. Liko Slingbar 350

B. Liko Slingbar 360

C. Liko Twinbar 670 / Slingbar Wide 670

D. Liko Sidebar Standard

Reason

Hill-Rom has become aware of a potential safety issue related to the Slingbar 350, Slingbar 360, Slingbar wide 670, and Sidebar standard, which could be attached to various Liko overhead and mobile lifts. A serious injury to a patient, caregiver, or bystander could result from tripping or otherwise falling in to the Sling bar, typically while the lift is parked or stored when not in use. The Sling hook on the end of the Sling bar could pierce a body structure, such as the head, shoulder, or groin. The affected sling bar designs have not been sold or distributed since May 2008.

Affected products

A. Liko Slingbar 350

Lot or serial number

Not applicable

Model or catalog number
  • 3156002
  • 3156004
  • 3156008
  • 3156015
Companies
Manufacturer

Hill-Rom Company, Inc.

1069 State Route 46 East

Batesville

47006-9167

Indiana

UNITED STATES



B. Liko Slingbar 360

Lot or serial number

Not applicable

Model or catalog number
  • 3156013
Companies
Manufacturer

Hill-Rom Company, Inc.

1069 State Route 46 East

Batesville

47006-9167

Indiana

UNITED STATES



C. Liko Twinbar 670 / Slingbar Wide 670

Lot or serial number

Not applicable

Model or catalog number
  • 3156012
  • 3156017
Companies
Manufacturer

Hill-Rom Company, Inc.

1069 State Route 46 East

Batesville

47006-9167

Indiana

UNITED STATES



D. Liko Sidebar Standard

Lot or serial number

Not applicable

Model or catalog number
  • 3156011
Companies
Manufacturer

Hill-Rom Company, Inc.

1069 State Route 46 East

Batesville

47006-9167

Indiana

UNITED STATES