Health product recall

LightSpeed VCT CT Scanner System

Starting date:
June 1, 2017
Posting date:
June 13, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63602

Reason

GE Healthcare has recently become aware of a potential safety issue that can result in a scan starting before the prescribed delay has completed. More specifically, if an exam is interrupted during the prep delay period using the emergency stop button and the scan is then resumed, the delay period may be shorter than expected. If personnel are in the scan room at the time of the premature scan initiation they can be exposed to scatter x-ray radiation.

Affected products

LightSpeed VCT CT Scanner System

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

B7864DC

Companies

Manufacturer

GE Medical Systems, LLC

3000 North Grandview Blvd.

Waukesha

53188

Wisconsin

UNITED STATES