LIFEPAK 15 V4 SpCO/SpMet Sensors
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
LNCS-II Rainbow DCIP 8 SpCO |
23JBP |
4068 |
LNCS-II Rainbow DCI 8 SpCO |
23HPT |
4067 |
LNCS-II Rainbow DCI 8 SpCO |
23HER |
4067 |
LNCS-II Rainbow DCI 8 SpCO |
23JTA |
4067 |
LNCS-II Rainbow DCI 8 SpCO |
23GJQ |
4067 |
LNCS-II Rainbow DCI 8 SpCO |
23JBN |
4067 |
RD Rainbow SpCO |
23JAU |
4036 |
Issue
Stryker received complaints indicating that the LIFEPAK 15 V4 displayed an error where “SpO2: Sensor does not support SpCO or SpMET.” The error results in preventing users from utilizing LIFEPAK 15 V4 SpCO and SpMet functionalities when connected to specific lots of Masimo Rainbow Sensors.
The specific sensor lots are compatible with the LIFEPAK 15 V1 and V2 devices.