LIFECODES LSA Class II
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
LIFECODES LSA Class II |
3016850 |
265200IVD |
LIFECODES LSA Class II |
3016697 |
265200IVD |
LIFECODES LSA Class II |
3016317 |
265200IVD |
LIFECODES LSA Class II |
3016532 |
265200IVD |
LIFECODES LSA Class II |
3016100 |
265200IVD |
Issue
Werfen has identified that two beads within the LIFECODES LSA class II assay kit(s) may present different DQB1 specificities than previously understood:
• Bead 42 / Antigen ID 242
Previously associated with DQB1*02:01, this bead instead presents DQB1*02:02
• Bead 44 / Antigen ID 244
Previously associated with DQB1*02:02, this bead instead presents DQB1*02:01
• Note: DQA1*02:01 is present on both beads and therefore not impacted.
Due to this issue, certain reactivity patterns may appear differently than expected.
Additional information
Details
| Immucor Gti Diagnostics, Inc. |
| 20925 Crossroads Circle, Waukesha, Wisconsin, United States, 53186 |
Get notified