This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Liaison® Testosterone XT

Brand(s)
Last updated

Summary

Product
Liaison® Testosterone XT
Issue
Medical devices - Performance
What to do

Users instructed to stop using affected devices. Destroy remaining affected inventory.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Liaison® Testosterone XT

136981

318410

Issue

The affected kit lots of Liaison® Testosterone XT may produce falsely low results in samples below 2 ng/ml.   Median values below 2 ng/ml show a negative bias to the median values listed in the IFU. The negative bias in samples below 2 ng/ml may impact interpretation of results relative to the reference range claims for the adult female populations and pediatric populations.

Recall start date: July 29. 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Clinical toxicology
Companies
Diasorin Inc.
1951 Northwestern Avenue, Stillwater, Minnesota, United States, 55082
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75961