LIAISON XL ANALYZER (2015-07-21)

Starting date:
July 21, 2015
Posting date:
August 11, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54570

Affected Products

A. LIAISON XL ANALYZER

Reason

Manufacturer became aware of the following: the "calibration expiry date is 9999-12-30 (never expires) could result in incorrect handling of incoming control. Frequent pop-up: "the period selected has an SD value of 0". Several periods may open in parallel for the same control. Error codes such as "pump warning 10" and "pump warning 14" appear during initialization of the instrument.

Affected products

A. LIAISON XL ANALYZER

Lot or serial number

>10 NUMBERS, CONTACT MFG

Model or catalog number

I0050