LIAISON XL ANALYZER (2015-07-21)
- Starting date:
- July 21, 2015
- Posting date:
- August 11, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54570
Affected Products
A. LIAISON XL ANALYZER
Reason
Manufacturer became aware of the following: the "calibration expiry date is 9999-12-30 (never expires) could result in incorrect handling of incoming control. Frequent pop-up: "the period selected has an SD value of 0". Several periods may open in parallel for the same control. Error codes such as "pump warning 10" and "pump warning 14" appear during initialization of the instrument.
Affected products
A. LIAISON XL ANALYZER
Lot or serial number
>10 NUMBERS, CONTACT MFG
Model or catalog number
I0050