LIAISON MDX/Integrated Cycler (2020-02-18)

Starting date:
February 18, 2020
Posting date:
February 28, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72449



Last updated: 2020-03-06

Affected Products

LIAISON MDX/Integrated Cycler

Reason

Magnetic field produced by the magnets (Platen and Consumable Cover) in the LIAISON MDX/Integrated Cycler may exceed 0.5 mT (5 Gauss). Workers with implanted ferromagnetic or electronic medical devices should not be exposed to static magnetic fields exceeding 0.5 mT (5 Gauss).

Affected products

LIAISON MDX/Integrated Cycler

Lot or serial number

All lots.

Model or catalog number

MOL1001

Companies
Manufacturer

Diasorin Inc.

1951 Northwestern Avenue, P.O. Box 285

Stillwater

55082

Minnesota

UNITED STATES