LIAISON MDX/Integrated Cycler (2020-02-18)
- Starting date:
- February 18, 2020
- Posting date:
- February 28, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72449
Last updated: 2020-03-06
Affected Products
LIAISON MDX/Integrated Cycler
Reason
Magnetic field produced by the magnets (Platen and Consumable Cover) in the LIAISON MDX/Integrated Cycler may exceed 0.5 mT (5 Gauss). Workers with implanted ferromagnetic or electronic medical devices should not be exposed to static magnetic fields exceeding 0.5 mT (5 Gauss).
Affected products
LIAISON MDX/Integrated Cycler
Lot or serial number
All lots.
Model or catalog number
MOL1001