This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

LIAISON C-Peptid

Starting date:
February 28, 2017
Posting date:
March 24, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62796

Reason

Recent internal testing highlighted the cross reactivity of LIAISON C-Peptid assay with human proinsulin which, in some rare cases, may result in an apparent increased c-peptide concentration in affected sample(s) tested using the LIAISON C-Peptid assay.

Affected products

LIAISON C-Peptid

Lot or serial number

All lots

Model or catalog number

316171

Companies

Manufacturer
Diasorin Inc.
1951 Northwestern Avenue, P.O. Box 285
Stillwater
55082
Minnesota
UNITED STATES