LIAISON C-Peptid
- Starting date:
- February 28, 2017
- Posting date:
- March 24, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-62796
Reason
Recent internal testing highlighted the cross reactivity of LIAISON C-Peptid assay with human proinsulin which, in some rare cases, may result in an apparent increased c-peptide concentration in affected sample(s) tested using the LIAISON C-Peptid assay.
Affected products
LIAISON C-Peptid
Lot or serial number
All lots
Model or catalog number
316171
Companies
- Manufacturer
-
Diasorin Inc.
1951 Northwestern Avenue, P.O. Box 285
Stillwater
55082
Minnesota
UNITED STATES