LFS Flat Screen Support Arm Systems
Brand(s)
Last updated
Summary
Product
LFS Flat Screen Support Arm Systems
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
LFS Flat Screen Support Arm Systems |
Not applicable |
B2-610-12-1 |
Issue
The stopper block on the LFS bracket could be broken or missing. Without this stopper, the flat screen of the LFS Flat Screen System can rotate beyond the original angle limit and cause excessive force to the spring shaft of the LFS bracket, causing it to break and shear off.
Recall Start Date: October 31, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Dkk Dai-Ichi Shomei Co. Ltd.
32-26, Sakashita 1-Chome, Tokyo, Japan, 174-0043
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74794
Get notified