Health product recall

LFS Flat Screen Support Arm Systems

Last updated

Summary

Product
LFS Flat Screen Support Arm Systems
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

LFS Flat Screen Support Arm Systems

Not applicable

B2-610-12-1
B2-610-12
B2-610-13
LFS

Issue

The stopper block on the LFS bracket could be broken or missing. Without this stopper, the flat screen of the LFS Flat Screen System can rotate beyond the original angle limit and cause excessive force to the spring shaft of the LFS bracket, causing it to break and shear off.
 

Recall Start Date: October 31, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Dkk Dai-Ichi Shomei Co. Ltd.

32-26, Sakashita 1-Chome, Tokyo, Japan, 174-0043

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74794

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe