Health product recall

Level 1® Fast Flow Fluid Warmer and Level 1® NORMOFLO Irrigation Fluid Warmer

Last updated

Summary

Product
Level 1® Fast Flow Fluid Warmer and Level 1® NORMOFLO Irrigation Fluid Warmer
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Level 1 Fast Flow Fluid Warmer

Not applicable.

H-1129
H-1100

Level 1 NORMOFLO Irrigation Fluid Warmer

Not applicable.

H1100
H-1200-EN-115V

Issue

Smiths Medical is issuing this letter to notify customers of a potential issue related to the Level 1 Fast Flow Fluid Warmers and Level 1 NORMOFLO Irrigation Fluid Warmers:

1. The return tube assembly inside the warmers may exhibit stress cracking that can propagate over time and potentially result in leaking recirculating fluid. Leaking recirculating fluid may potentially damage other components within the warmer, such as the printed circuit board.

2. The cable inside the warmers that connects to the external on/off switch may potentially exhibit delamination of insulation which may propagate over time and may lead to corrosion and a break in the electrical circuit.

Recall Start Date: July 2, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Smiths Medical Asd, Inc.

6000 Nathan Lane N., Minneapolis, Minnesota, United States, 55442

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75836

Get notified

Receive emails about new and updated recall and safety alerts.