LetsGetChecked Blood & Urine Sample Collection Kit Device (2019-07-11)
- Starting date:
- July 11, 2019
- Posting date:
- August 9, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70669
Last updated: 2019-08-09
Affected Products
A. LetsGetChecked Blood Sample Collection Kit Device
B. LetsGetChecked Blood & Urine Sample Collection Kit
Reason
The kit contains BD microtainer tubes subject to recall. The tubes are being recalled as there is the potential for blood leakage due to a hole or damage at the made of the tube.This defect could result in blood exposure from the leaked sample or potential erroneous results due to an inappropriate blood-to-additive ratio. There is also potential for tube underfill due to another type of tube damage where the blood does not leak, and therefore may not be detected by the user. This could lead to potential underfill with an incorrect blood-to-additive ratio and potential erroneous results if the sample is tested.
Affected products
A. LetsGetChecked Blood Sample Collection Kit Device
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
LGC-BC-KIT
Companies
- Manufacturer
-
PRIVAPATH DIAGNOSTICS LTD.
UNIT 3, ADELPHI HOUSE, GEORGES STREET UPPER
DUN LAOGHAIRE
IRELAND
B. LetsGetChecked Blood & Urine Sample Collection Kit
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
LGC-BUC-KIT
Companies
- Manufacturer
-
PRIVAPATH DIAGNOSTICS LTD.
UNIT 3, ADELPHI HOUSE, GEORGES STREET UPPER
DUN LAOGHAIRE
IRELAND