Health product recall

Leica Cryostats

Last updated

Summary

Product
Leica Cryostats
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Leica Cryostats More than 10 numbers, contact manufacturer. CM1950 CM1900 CM3050S CM1850

Issue

On August 6, 2024, Leica Biosystems became aware, through post market surveillance, of a customer using flammable freezing spray within the cryostat chamber. The air/spray mixture in the chamber ignited, leading to an explosion causing slight burns to the arms and face of the technician. The scope of the recall has been expanded to include all devices produced prior to 2009/01/01.

Recall start date: August 29, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Pathology
Companies

Leica Biosystems Nussloch GmbH

Heidelberger Street 17-19, Nussloch, Germany, 69226

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76011

Get notified

Receive notifications for new and updated recalls and alerts by category.