This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Laparoscopic Manual Instruments and Accessories (2014-11-21)

Starting date:
November 21, 2014
Posting date:
November 28, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42697

Recalled Products

Laparoscopic Manual Instruments and Accessories

Reason

The parameters provided in the Laparoscopic Manual Instruments instructions for use (IFU 1000-401-070 revision G or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.

Affected products

Laparoscopic Manual Instruments and Accessories

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

Not applicable

Companies
Manufacturer

Stryker Endoscopy

5900 Optical Court

San Jose

95138

California

UNITED STATES