LAERDAL SUCTION UNIT, VARIOUS COMPONENTS

Starting date:
August 15, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60044

Affected Products

A. LAERDAL SUCTION UNIT W/REUSABLE CANNISTER

B. LAERDAL SUCTION UNIT W/SERRES SUCTION BAG

C. LAERDAL SUCTION UNIT NIMH BATTERY

Reason

AWARE OF A PROBLEM RELATED TO USE OF THE LAERDAL SUCTION UNIT (LSU) WHEN USED WITH THE LSU BATTERY THAT HAS BEEN STORED OR INSTALLED A LOW TEMPERATURES FOR A PROLONGED PERIOD OF TIME. THE LSU MAY SHUT OFF AFTER SOME SECONDS WHEN OPERATED AT 350 MMHG OR 500+ MMHG SETTINGS.

Affected products

A. LAERDAL SUCTION UNIT W/REUSABLE CANNISTER

Lot or serial number

Not applicable

Model or catalog number

78000016

78000020



 

Companies
Manufacturer

Laerdal Medical AS

Tanke Svilandsgate 30, P.O. Box 377,

Stavanger, Rogaland Fylke

4002

NORWAY



B. LAERDAL SUCTION UNIT W/SERRES SUCTION BAG

Lot or serial number

Not applicable

Model or catalog number

78003016

78003020

Companies
Manufacturer

Laerdal Medical AS

Tanke Svilandsgate 30, P.O. Box 377,

Stavanger, Rogaland Fylke

4002

NORWAY



C. LAERDAL SUCTION UNIT NIMH BATTERY

Lot or serial number

Not applicable

Model or catalog number

780800

Companies
Manufacturer

Laerdal Medical AS

Tanke Svilandsgate 30, P.O. Box 377,

Stavanger, Rogaland Fylke

4002

NORWAY