Lactate Dehydrogenase L-P (LDH) (NAD) (2020-08-15)
- Starting date:
- August 15, 2020
- Posting date:
- August 28, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73799
Last updated:
Affected Products
Lactate Dehydrogenase L-P (LDH) (NAD)
Reason
Randox have released an update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate dehydrogenase assay on Rx instruments. Additional pipette washes can be implemented as described in the technical bulletin.
Affected products
Lactate Dehydrogenase L-P (LDH) (NAD)
Lot or serial number
- 523902
- 536342
Model or catalog number
LD3842
Companies
- Manufacturer
-
Randox Laboratories Ltd.
55 Diamond Road
Crumlin, County Antrim
BT29 4QY
UNITED KINGDOM