This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

KleenSpec Corded Illuminator (2015-10-09)

Starting date:
October 9, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55630

Affected Products

KleenSpec Corded Illuminator (2015-10-09)

Reason

Welch Allyn has been made aware that the KleenSpec 788 Corded Illuminator has the potential to be incorrectly connected to a commonly available AC power cord instead of the illuminator low voltage transformer power cord which is supplied with the illuminator. This may result in a shattered lamp bulb and a potential electric shock hazard. Commonly available AC line power cords do not have an integrated wall transformer and therefore, the electrical current from the outlet is fully transferred into the device. There have been no associated injuries, but the misconnection can be reproduced.  For example, the illuminator cord connector may fit into AC power cords from common consumer equipment, such as laptop computers and other medical devices that have similar power cord connector.

Affected products

KleenSpec Corded Illuminator (2015-10-09)

Lot or serial number

All lots.

Model or catalog number

73410

78800

78810

Companies
Manufacturer

Welch Allyn Inc.

4341 State Street Road

Skaneateles Falls

13153-0220

New York

UNITED STATES