Health product recall

Kiwi® Vacuum Assisted Delivery System

Last updated

Summary

Product
Kiwi® Vacuum Assisted Delivery System
Issue
Medical devices - Performance
What to do

Manufacturer will contact custmers with instructions on return of affected products. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Kiwi® Vacuum Assisted Delivery System

Contact the manufacturer.

VAC-6000MT

Issue

The device may break at the Traction Force Indicator (TFI) joint during use when traction is applied at approximately a 45-degree angle.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Obstetrics and gynaecology
Companies

Clinical Innovations, LLC

747 West 4170 South, Murray, Utah, United States, 84123

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81826