KANGAROO CONNECT ENTERAL FEEDING PUMP (2016-04-04)

Starting date:
April 4, 2016
Posting date:
April 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58010

Affected products 

A. KANGAROO CONNECT ENTERAL FEEDING PUMP

Reason

This voluntary recall is being conducted following reports that the occlusion alarm failed to sound as intended due to a software anomaly. As a result of this software anomaly, the pump may fail to alarm when the pump is restarted or the cassette is reloaded

Affected products

A. KANGAROO CONNECT ENTERAL FEEDING PUMP

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

384400

Companies
Manufacturer

Covidien LLC

15 Hampshire Street

Mansfield

02048

Massachusetts

UNITED STATES