KANGAROO CONNECT ENTERAL FEEDING PUMP (2016-04-04)
- Starting date:
- April 4, 2016
- Posting date:
- April 22, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-58010
Affected products
A. KANGAROO CONNECT ENTERAL FEEDING PUMP
Reason
This voluntary recall is being conducted following reports that the occlusion alarm failed to sound as intended due to a software anomaly. As a result of this software anomaly, the pump may fail to alarm when the pump is restarted or the cassette is reloaded
Affected products
A. KANGAROO CONNECT ENTERAL FEEDING PUMP
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
384400
Companies
- Manufacturer
-
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES