Juno DRF (2019-11-14)
- Starting date:
- November 14, 2019
- Posting date:
- November 29, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71746
Last updated: 2019-12-02
Affected Products
Juno DRF
Reason
It was found that the angulation chain is very close to the design limit for metal fatigue and, under severe workload conditions might fail before the expected lifetime of 10 years.
Affected products
Juno DRF
Lot or serial number
Not applicable.
Model or catalog number
9784729816
9784729831
Companies
- Manufacturer
-
Villa Sistemi Medicali S.P.A.
via Delle Azalee 3
Buccinasco
20090
ITALY