Health product recall

Juno DRF (2019-11-14)

Starting date:
November 14, 2019
Posting date:
November 29, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71746



Last updated: 2019-12-02

Affected Products

Juno DRF

Reason

It was found that the angulation chain is very close to the design limit for metal fatigue and, under severe workload conditions might fail before the expected lifetime of 10 years.

Affected products

Juno DRF

Lot or serial number

Not applicable.

Model or catalog number

9784729816

9784729831

Companies
Manufacturer

Villa Sistemi Medicali S.P.A.

via Delle Azalee 3

Buccinasco

20090

ITALY