This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Juno DRF

Starting date:
December 13, 2016
Posting date:
January 6, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61706

Reason

A patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and the table was moving from vertical position.

Affected products

Juno DRF

Lot or serial number

All lots

Model or catalog number

9784729831

Companies

Manufacturer

Villa Sistemi Medicali S.P.A.

via Delle Azalee 3

Buccinasco

20090

ITALY