JuggerLoc Disposable Cannula Kit - Drill Guide
Brand(s)
Last updated
Summary
Product
JuggerLoc Disposable Cannula Kit - Drill Guide
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
JuggerLoc Disposable Cannula Kit - Drill Guide | 0002467342 0002476076 0002497715 0002504035 0002446333 0002493107 0002496066 0002446326 | 110027357 |
Issue
Zimmer Biomet is conducting a lot specific medical device recall for the JuggerLoc Disposable Cannula Kit - Drill Guide. Zimmer Biomet has received multiple complaints where it was reported that the guides were discernibly warped when pulled out of packaging. Following investigation, the warping has been attributed to the cleaning process used during supplier manufacturing.
Recall Start Date: October 28, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies
Biomet Sports Medicine
56 East Bell Drive, Warsaw, Indiana, United States, 46581
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76441
Get notified
Receive notifications for new and updated recalls and alerts by category.