JuggerLoc Disposable Cannula Kit - Drill Guide
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
JuggerLoc Disposable Cannula Kit - Drill Guide |
0002467342 |
110027357 |
Issue
Zimmer Biomet is conducting a lot specific medical device recall for the JuggerLoc Disposable Cannula Kit - Drill Guide. Zimmer Biomet has received multiple complaints where it was reported that the guides were discernibly warped when pulled out of packaging. Following investigation, the warping has been attributed to the cleaning process used during supplier manufacturing.
Recall Start Date: October 28, 2024
Additional information
Details
Biomet Sports Medicine
56 East Bell Drive, Warsaw, Indiana, United States, 46581
Get notified