This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Joerns Healthcare EasyCare Bed (2014-10-07)

Starting date:
October 7, 2014
Posting date:
October 29, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41927

Recalled Products

Joerns Healthcare EasyCare Bed

Reason

Joerns healthcare has identified an issue with EasyCare Beds configured w/the care-lock system where in some instances the care-lock system, after being engaged by stepping on the red or right side pedal, through normal daily use and patient/care-giver interaction with the bed can disengage i.e., revert to a non-locked state. The care-lock system is located at the foot end of the bed, there are also steering locks located on the dual wheel casters at the head of the bed. This does not affect the steering locks located on the dual wheel casters at the head of the bed.

Affected products

Joerns Healthcare EasyCare Bed

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

EASYCARE

Companies
Manufacturer

Joerns Healthcare,

Avenida Pedro Hinojosa No. 26, Ciudad Industrial,

Matamoros Tamps

87499

MEXICO