Health product recall

Jackson-Pratt 3-Spring Reservoir Kits

Brand(s)
Last updated

Summary

Product
Jackson-Pratt 3-Spring Reservoir Kits
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Jackson-Prat Universal Connector Pvc Tubing Anti Reflux Transparent Y

210751D6

SU130-4601

Jackson-Pratt 3-Spring Reservoir With Silicone Adapters, 400ml

231251D5

SU130-475

Jackson-Pratt 3-Spring Reservoir Kits

More than 10 numbers, contact manufacturer.

SU130-403D
SU130-401D
SU130-402D

Issue

Cardinal Health is issuing a lot number specific voluntary removal of Jackson-Pratt™ 3-Spring Reservoir Kits due to specific products being shipped to users before undergoing sterilization. A review of shipments revealed that specific Jackson-Pratt™ 3-Spring Reservoir Kits were not shipped to a sterilizer prior to product release. A third-party carrier inadvertently directly shipped to a Cardinal Health distribution center rather than the third-party sterilizer. Cardinal Health is issuing an urgent medical device product recall to request the return of all affected product.

Recall start date: March 28, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Cardinal Health
3651 Birchwood Drive, Waukegan, Illinois, United States, 60085
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75373