Health product recall

iVIEW DAB, ultraView Universal DAB, ultraView SISH DNP, OptiView DAB IHC Detection Kit and OptiView Amplification Kit, CINtec PLUS Cytology Kit, Hematoxylin II (2017-12-27)

Starting date:
December 27, 2017
Posting date:
January 12, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65668

Affected products

  1. iVIEW DAB Detection Kit
  2. ultraView Universal DAB Detection Kit
  3. ultraView SISH DNP Detection Kit
  4. OptiView DAB IHC Detection Kit
  5. OptiView Amplification Kit
  6. CINtec PLUS Cytology Kit
  7. Hematoxylin II

Reason

Related to customer complaint investigation, a potential situation concerning dispenser issues with Hematoxylin II and Horseradish peroxidase reagents where it was reported that these reagents dispensers could be leaking or sticking. This failure could result in a complete or partial dispense failure of a reagent critical to the staining reaction, potentially leading to erroneous results.

Affected products

A. iVIEW DAB Detection Kit

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

760-091

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



B. ultraView Universal DAB Detection Kit

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

760-500

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



C. ultraView SISH DNP Detection Kit

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

790-2208

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



D. OptiView DAB IHC Detection Kit

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

800-098

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



E. OptiView Amplification Kit

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

760-700

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



F. CINtec PLUS Cytology Kit

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 760-099
  • 860-099
Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



G. Hematoxylin II

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

605-100

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY