IVEA MODEL 500A (2017-10-11)

Starting date:
October 11, 2017
Posting date:
October 27, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64954

Affected products

IVEA MODEL 500A

Reason

Rotational forces on the 16082 dowel pin during deploying or folding can cause the pin to slowly "screw" its way out of position. If the pin migrates enough to allow the brace to detach from the leg pivot fitting while a patient is ambulating with weight on the handles, the product could slump to one side. This may cause the patient to stumble.

Affected products

IVEA MODEL 500A

Lot or serial number
  • 100325,
  • 100463,
  • 100476,
  • 100478,
  • 100482,
  • 100486,
  • 100487,
  • 100491,
  • 100556,
  • 100569,
  • 100570,
  • 100571,
  • 100572,
  • 100574,
  • 100592,
  • 100598,
  • 100600,
  • 100602,
  • 100605,
  • 100615,
  • 100620,
  • 100626,
  • 100627,
  • 100640,
  • 100641,
  • 100642,
  • 100643,
  • 100645,
  • 100646,
  • 100647,
  • 100650,
  • 100677
Model or catalog number

500A

Companies
Manufacturer

FIREFLY MEDICAL INC.,

320 EAST VINE DRIVE

FORT COLLINS

80524

Colorado

UNITED STATES