Health product recall

ITS4850A IntelliVue Telemetry System - MONITOR (2018-11-11)

Starting date:
November 11, 2018
Posting date:
November 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68424

Affected products

ITS4850A IntelliVue Telemetry System - MONITOR

Reason

The volume of the MX40 speaker in these devices may be diminished or not audible above 4,500 feet when operating in monitor mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the power on self-test.

Affected products

ITS4850A IntelliVue Telemetry System - MONITOR

Lot or serial number

Not applicable

Model or catalog number
  • MX40
Companies
Manufacturer

Philips Medical Systems

3000 Minuteman Road

Andover

01810

UNITED STATES