ITS4850A Intellivue Telemetry System-Main Unit
- Starting date:
- March 6, 2017
- Posting date:
- March 31, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-62852
Reason
Philips has recently discovered that under specific 802.11 network conditions, a partial re-association of the MX40 WLAN monitor to a compatible central monitoring system ("information center") is possible. In this state, although the information center provides a visible and audible "no data tele" INOP alert, the MX40 WLAN itself enters telemetry mode, i.e., its screen turns off in one minute and local alarming is disabled.
Affected products
ITS4850A Intellivue Telemetry System-Main Unit
Lot or serial number
More than 100 numbers, contact manufacturer
Model or catalog number
ITS4850A
Companies
- Manufacturer
-
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES