Health product recall

IRMA TRUpoint Cartridges (2015-09-16)

Starting date:
September 16, 2015
Posting date:
October 22, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55522

Affected Products

  • A. IRMA TRUpoint CC Cartridges
  • B. IRMA TRU point BG (Blood Gas) Cartridges

Reason

The affected lots may report values outside the performance specifications for pH. After a certain point in shelf life, the pH sensor may give results that are falsely elevated.

Affected products

A. IRMA TRUpoint CC Cartridges

Lot or serial number

ASBEZ

ASCXE

ASEMB

ASHWQ

ASJMY

ASLGI

ASMXR

ASOQO

ASRXC

ASTQR

ASWZI

Model or catalog number

39903

Companies
Manufacturer

LifeHealth, LLC

2656 PATTON ROAD

Roseville

55113

Minnesota

UNITED STATES



B. IRMA TRU point BG (Blood Gas) Cartridges

Lot or serial number

EOSCU

EOTUZ

EOVMW

EOYUL

EPANW

EPCED

EPDVM

EPFOJ

EPIXX

EPKOJ

EPNXA

Model or catalog number

048103

Companies
Manufacturer

LifeHealth, LLC

2656 PATTON ROAD

Roseville

55113

Minnesota

UNITED STATES